Customization: | Available |
---|---|
After-sales Service: | 1 year warranty lifelong free technical support |
Warranty: | 1 year warranty lifelong free technical support |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specification name | Technical specification parameters |
Scope of application | This product is suitable for wastewater with total iron in the range of 0~5mg/l. |
Test Method | o-Phenanthroline colourimetric method |
Measurement range | 0~1~5mg/L, expandable |
Lower limit of detection | 0.02 |
Resolution | 0.001 |
Accuracy | ±10% |
Repeatability | ±5%F.S |
Zero drift | ±5%F.S |
Range Drift | ±5%F.S |
Measurement Period | Minimum test cycle 30min, settable |
Sampling Cycle | Time interval (adjustable), whole point or trigger measurement mode, settable |
Calibration Cycle | Automatic calibration (1~99 days adjustable), according to the actual water samples, manual calibration can be set up |
Maintenance Cycle | Maintenance interval greater than 1 month, each time about 30min |
Human-machine operation | Touch screen display and command input |
Self-test protection | Self-diagnosis of instrument working status, no data loss after abnormal or power failure After abnormal reset or power failure, the instrument automatically eliminates residual reactants and resumes work automatically. |
Data storage | Not less than one year data storage |
Input Interface | Switch quantity |
Output Interface | 1 RS232 output, 1 RS485 output, 2 4~20mA outputs |
Working environment | Indoor work, the recommended temperature 5 ~ 28 ºC, humidity ≤ 90% (no condensation) |
Power supply and power consumption | AC230±10%V, 50~60Hz, 5A |
Dimension | H1500×W550×D450 (mm) |