Customization: | Available |
---|---|
After-sales Service: | 1 year warranty lifelong free technical support |
Warranty: | 1 year warranty lifelong free technical support |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
No. | Item | Technical Parameters |
1 | Application Range | This method is suitable for waste water of nickel in the range of 0~30 mg/L. |
2 | Test Methods | The measure way of nickel:the dimethylglyoxime spectrophotometry |
3 | Lower limit of Detection | 0~30mg/L |
4 | Lower limit of Detection | 0.05 |
5 | DPI | 0.001 |
6 | Accuracy | ±10% or ± 0.1mg/L (the bigger one between them) |
7 | Repeatability | 10% or 0.1mg/L (the bigger one between them) |
8 | Zero Drift | ±1 |
9 | Span Drift | ±10% |
10 | Measurement cycle | Minimum 20min, it can be set |
11 | Sampling period | Time interval (adjustable), integral hour or trigger measurement mode can be set. |
12 | Calibration cycle | Time interval (adjustable), integral hour or trigger measurement mode can be set. |
13 | Maintenance cycle | Maintenance interval is more than one month, about 30 minutes each time. |
14 | Man-machine operation | Touch screen display and instruction input. |
15 |
Self checking protection | Working status is self-diagnostic, abnormal or power failure will not lose data. Automatically eliminates residual reactants and resumes work after abnormal reset or power failure. |
16 | Data storage | No less than half a year data storage |
17 | Input interface | Switch quantity |
18 | Output interface | Two RS232 digital output, One 4-20mA analog output |
19 | Working conditions | Working indoors; temperature 5-28ºC; relative humidity≤90% (no condensation,no dew) |
20 | Power Supply and Consumption | AC230±10%V, 50~60Hz, 5A |
21 | Size | Height 1500×Width 550×Depth 450(mm) |